Health product recall

Multidiagnost Eleva - Flat Detector - Image Detector, Allura XPER FD10/10 - Main Unit,Allura XPER FD20/20-Main Unit,Allura XPER FD10 and Allura XPER FD20 System - Main Unit (2014-09-11)

Starting date:
September 11, 2014
Posting date:
October 1, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-41563

Recalled Products

A) Multidiagnost Eleva - Flat Detector - Image Detector

B) Allura XPER FD10/10 - Main Unit

C) Allura XPER FD20/20-Main Unit

D) Allura XPER FD10

E) Allura XPER FD20 System - Main Unit

Reason

Philips has become aware that if a footswitch is frequently used on an anti-fatigue mat, on a not flat surface or in the pedestal, the footswitch pedals may get bent. This might cause an intermittent or continuous inability of making live fluoroscopy images or exposures.

Affected products

A) Multidiagnost Eleva - Flat Detector - Image Detector

 

Lot or serial number

107

126

13

194

215

65

69

78

 

Model or catalog number

0708MD02

 

Companies
Manufacturer

Philips Medical Systems Nederland B.V.

Veenpluis 4-6, 5684 PC

Best, Netherlands



B) Allura XPER FD10/10 - Main Unit

 

Lot or serial number

215

 

Model or catalog number

ALLURA XPER FD10/10

 

Companies
Manufacturer

Philips Medical Systems Nederland B.V.

Veenpluis 4-6, 5684 PC

Best, Netherlands



C) Allura XPER FD20/20-Main Unit

 

Lot or serial number

527

567

607

Model or catalog number

ALLURA XPER FD20/20

 

Companies
Manufacturer

Philips Medical Systems Nederland B.V.

Veenpluis 4-6, 5684 PC

Best, Netherlands



D) Allura XPER FD10

 

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

ALLURA XPER FD10

 

Companies
Manufacturer

Philips Medical Systems Nederland B.V.

Veenpluis 4-6, 5684 PC

Best, Netherlands



E) Allura XPER FD20 System - Main Unit

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

ALLURA XPER FD20

Companies
Manufacturer

Philips Medical Systems Nederland B.V.

Veenpluis 4-6, 5684 PC

Best, Netherlands