Health product recall

Mueller-Hinton Broth with 3% lysed horse blood (2018-03-18)

Starting date:
March 18, 2018
Posting date:
November 28, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-68384

Affected products

Mueller-Hinton Broth with 3% lysed horse blood

Reason

Beckman coulter has received customer complaints of lowered minimum inhibitory concentration (mic) for antimicrobial agent’s Cefaclor (CFR) and Penicillin (P) with control organism streptococcus pneumoniae (ATCC 49619) during Quality Control (QC) testing on MICroSTREP plus Panel Type 1 hydrated with the broth lot mentioned above.

The performance issue may be detected by QC. However, if the broth is put into use and QC testing is not performed, or the panels pass QC testing and are used for clinical isolate testing, then there is the potential for false susceptible results due the organism not growing. This may result in progression of infection or failure to control a bacterial infection leading to other complications.

Affected products

Mueller-Hinton Broth with 3% lysed horse blood

Lot or serial number

1005986140

Model or catalog number

B1015-25

Companies
Manufacturer
Beckman Coulter Inc.
250 S. Kraemer Blvd.
Brea
92821
UNITED STATES