Health product recall

MSK Extreme and Optima MR430S 1.5T MRI Systems (2014-12-29)

Starting date:
December 29, 2014
Posting date:
January 7, 2015
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-43103

Recalled Products

A) MSK Extreme - MRI System

B) Optima MR430S 1.5T - MRI System

Reason

Magnet Rundown Unit (MRU) may not be connected to the magnet. If the MRU is disconnected, the MRU will not function as expected and will not quench the magnet when the button is pressed. In emergency situations, a disconnected MRU could delay removal of a ferrous object from the magnet, potentially resulting in life-threatening injuries. The MRU must be connected to the magnet at all times.

Affected products

A) MSK Extreme - MRI System

Lot or serial number

905521ONI

Model or catalog number

AB2000

Companies
Manufacturer

GE Medical Systems, LLC

301 Ballardvale St., Suite 4,

Wilmington

01887

Massachusetts

UNITED STATES



B) Optima MR430S 1.5T - MRI System

Lot or serial number

514697ONI

Model or catalog number

OPTIMA MR430S 1.5T (5000-0002)

Companies
Manufacturer

GE Medical Systems, LLC

301 Ballardvale St., Suite 4,

Wilmington

01887

Massachusetts

UNITED STATES