Health product recall

Monnal T60 Ventilator (2019-09-30)

Starting date:
September 30, 2019
Posting date:
October 18, 2019
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-71309


Last updated:

Affected Products

Monnal T60 Ventilator

Reason

2 Issues being addressed:

1. Early hardening of the silicon membrane after the sterilization cycle which is dependent on the disinfectant solution used during the immersion process. This change in the characteristics of the silicon membrane could have an impact on the ventilation. 

2. Air Liquide Medical Systems wishes to better specify the need for filters both to protect the integrity of the device and to prevent cross-contamination.

Affected products

Monnal T60 Ventilator

Lot or serial number

All serial numbers.

Model or catalog number

KA010000

KA017115

Companies
Manufacturer

Air Liquide Medical Systems

6 Rue Georges Besse, Parc de Haute Technologie

Antony

92182

FRANCE