Monnal T60 Ventilator (2019-09-30)
- Starting date:
- September 30, 2019
- Posting date:
- October 18, 2019
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-71309
Last updated: 2019-10-18
Affected Products
Monnal T60 Ventilator
Reason
2 Issues being addressed:
1. Early hardening of the silicon membrane after the sterilization cycle which is dependent on the disinfectant solution used during the immersion process. This change in the characteristics of the silicon membrane could have an impact on the ventilation.
2. Air Liquide Medical Systems wishes to better specify the need for filters both to protect the integrity of the device and to prevent cross-contamination.
Affected products
Monnal T60 Ventilator
Lot or serial number
All serial numbers.
Model or catalog number
KA010000
KA017115
Companies
- Manufacturer
-
Air Liquide Medical Systems
6 Rue Georges Besse, Parc de Haute Technologie
Antony
92182
FRANCE