Monaco RTP System (2019-07-01)
- Starting date:
- July 1, 2019
- Posting date:
- July 19, 2019
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-70457
Last updated: 2019-07-19
Affected Products
Monaco RTP System
Reason
Monitor units can be incorrect when following a specific workflow in MONACO. When creating 3D plans using either MU or dose weighting modes, if the user changes the number of fractions, rescales the plan and then changes the wedge angle, the monitor units are scaled incorrectly.
Affected products
Monaco RTP System
Lot or serial number
More than 10 numbers, contact manufacturer.
Model or catalog number
MONACO RTP SYSTEM
Companies
- Manufacturer
-
Elekta INC.
13723 RIVERPORT DRIVE,, SUITE 100
Maryland Heights
63043
Missouri
UNITED STATES