This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

MONACO 5 (2016-05-18)

Starting date:
May 18, 2016
Posting date:
May 30, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-58620

Affected products 

A. MONACO 5

Reason

When DICOM exporting a 3D Monaco plan and the "Composite Field Sequencing" (CFS) checkbox is selected, the Y jaws will snap to the port shape when they should remain where they were defined. This occurs when the port shape used to define or edit the MLC positions extends beyond or inside the actual jaw positions.

Affected products

A. MONACO 5

Lot or serial number

VERSIONS 5.10.00 AND HIGHER

Model or catalog number

MONACO RTP SYSTEM

Companies
Manufacturer

Elekta Business Area Software Systems Impac Medical Systems

Inc.

13723 Riverport Drive

Missouri

UNITED STATES