MONACO 5 (2016-05-18)
- Starting date:
- May 18, 2016
- Posting date:
- May 30, 2016
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-58620
Affected products
A. MONACO 5
Reason
When DICOM exporting a 3D Monaco plan and the "Composite Field Sequencing" (CFS) checkbox is selected, the Y jaws will snap to the port shape when they should remain where they were defined. This occurs when the port shape used to define or edit the MLC positions extends beyond or inside the actual jaw positions.
Affected products
A. MONACO 5
Lot or serial number
VERSIONS 5.10.00 AND HIGHER
Model or catalog number
MONACO RTP SYSTEM
Companies
- Manufacturer
-
Elekta Business Area Software Systems Impac Medical Systems
Inc.
13723 Riverport Drive
Missouri
UNITED STATES