Health product recall

MONACO 5

Starting date:
November 3, 2016
Posting date:
November 15, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-61054

Affected Products

Reason

When creating 3D plans using dose weighting mode, if the user rescales the plan, selects "reset", and then changes the wedge angle, the plan dose and monitor units (MU) should be returned to the original values before the rescale. Instead, the system correctly restores the edited wedge field but incorrectly applies the scale reset value again to all other beams in the plan.

Affected products

MONACO 5

Lot or serial number
  • Version 5.11
  • Version 5.20
Model or catalog number
  • MONACO RTP SYSTEM
Companies
Manufacturer
Elekta Business Area Software Systems Impac Medical Systems, Inc.
13723 RIVERPORT DRIVE SUITE 100
Maryland Heights
63043
Missouri
UNITED STATES