Health product recall

MONACO (2015-10-14)

Starting date:
October 14, 2015
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-55712

Affected Products

MONACO

Reason

When the 100% isodose value is edited on the isodose control, the dose normalization is automatically updated. If the user does not notice this update and then rescales to the relative isoline and does not read the dose normalization value displayed on the rescale bar, the rescale could be other than intended. When the user optimizes, then changes the bolus assigment of the beams, and then selects "calculate" (not "optimize" again), the bolus assignment to beams can get scrambled.

Affected products

MONACO

Lot or serial number

Version 5.10.00 and higher

Model or catalog number

MONACO

Companies
Manufacturer
Elekta Business Area Software Systems Impac Medical Systems, Inc.
13723 RIVERPORT DRIVE SUITE 100
Maryland Heights
63043
Missouri
UNITED STATES