This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

MONACO (2015-09-15)

Starting date:
September 15, 2015
Posting date:
October 5, 2015
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-55210

Affected Products

MONACO

Reason

Two issues affecting the Monaco device: 1) when CT images are created this latter way and imported in Monaco, the image is displayed correctly but the electron densities are incorrectly flipped in both the right-left and superior-inferior directions. 2) when using multiple prescription and the first prescription has forced densities, these forced densities are not being maintained with the second prescription.

Affected products

MONACO

Lot or serial number

Not applicable

Model or catalog number

MONACO

Companies
Manufacturer

Elekta Business Area Software Systems Impac Medical Systems

Inc.

13723 Riverport Drive

Missouri

UNITED STATES