This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Monaco (2014-12-19)

Starting date:
December 19, 2014
Posting date:
December 29, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-43037

Recalled Products   

Monaco 

Reason

In the case where CT slice thickness is not the same as CT slice spacing the image statistics tool will report an incorrect HU. If the user bases their CT to ED file on this incorrect density information, dose calculation errors will result. The magnitude of the dose calculation error will be dependent on the magnitude of the HU error.

Affected products

Monaco 

Lot or serial number

Not applicable  

Model or catalog number

MONACO

Companies
Manufacturer

Elekta Business Area Software Systems Impac Medical Systems Inc.

13723 Riverport Drive, Suite 100

Maryland Heights

63043

Missouri

UNITED STATES