This page has been archived on the Web
Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.
MONACO
- Starting date:
- November 17, 2016
- Posting date:
- December 5, 2016
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-61282
Reason
Contours are not handled correctly if the computed tomography (CT) dataset contains slices with DICOM Z coordinates of xx.x50mm. If the CT data contain slices that occur at positions of xx.x50mm, there can be portions of the volume where there is a mismatch between the contour graphics shown on the screen and the contour data used during the optimization and calculation of dose. Unloading and reloading the patient data correct the problem.
Affected products
MONACO
Lot or serial number
Version 5.0 and higher
Model or catalog number
MONACO
Companies
- Manufacturer
-
Elekta Business Area Software Systems Impac Medical Systems, Inc.
13723 RIVERPORT DRIVE
SUITE 100
Maryland Heights
63043
Missouri
UNITED STATES