- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-61282
Reason
Contours are not handled correctly if the computed tomography (CT) dataset contains slices with DICOM Z coordinates of xx.x50mm. If the CT data contain slices that occur at positions of xx.x50mm, there can be portions of the volume where there is a mismatch between the contour graphics shown on the screen and the contour data used during the optimization and calculation of dose. Unloading and reloading the patient data correct the problem.
Affected products
MONACO
Lot or serial number
Version 5.0 and higher
Model or catalog number
MONACO
Companies
- Manufacturer
-
Elekta Business Area Software Systems Impac Medical Systems, Inc.