MitraClip XTR Clip Delivery System (2019-04-30)
- Starting date:
- April 30, 2019
- Posting date:
- May 14, 2019
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type I
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-69938
Last updated:
Affected Products
MitraClip XTR Clip Delivery System
Reason
Abbott has confirmed reports of XTR Clips unexpectedly opening and becoming nonfunctional, resulting from unintended excessive force applied to the clip during implantation. Once the clip becomes nonfunctional, the inability to close and remove the device has led to surgery or additional intervention.
Affected products
MitraClip XTR Clip Delivery System
Lot or serial number
Not applicable.
Model or catalog number
CDS0602-XTR
Companies
- Manufacturer
-
Abbott Vascular
3200 Lakeside Drive
Santa Clara
95054-2807
California
UNITED STATES