MITRACLIP CLIP DELIVERY SYSTEM (2016-02-08)

Starting date:
February 8, 2016
Posting date:
February 22, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type I
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-57196

Affected Products

MITRACLIP CLIP DELIVERY SYSTEM

Reason

Abbott Vascular (AV) has recently received reports where the user attempted implanting a MitraClip, but the clip could not be detached from the delivery system due to a mandrel fracture. A mandrel fracture may occur if tension is present on the mandrel when turning the actuator knob to deploy the clip. Tension is present if the arm positioner is on the "closed" side of neutral, as opposed to being in the neutral position during clip deployment. The Field Safety Notice (FSN) is to alert users and provide an alternative deployment sequence that will ensure tension is not present in the system.

Affected products

MITRACLIP CLIP DELIVERY SYSTEM

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

CDS0201