Minicon Clinical Sample Concentrators (B-15 & CS-15) (2018-06-11)
- Starting date:
- June 11, 2018
- Posting date:
- June 29, 2018
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-67130
Affected products
Minicon Clinical Sample Concentrators (B-15 & CS-15)
Reason
Since the 22nd of February 2017, 39 Minicon IVD complaints have been received. 97% of the complaints received were for device failing to concentrate the samples within the internal specification of less than or equal to 3 hours. A trend report to the Irish Competent Authority (HPRA) was submitted on March 7th 2018, (HPRA ref v35094). On May 3rd 2018, the HPRA requested that a field safety notice be issued based on their opinion that there exists a potential risk to patient safety.
Affected products
Minicon Clinical Sample Concentrators (B-15 & CS-15)
Lot or serial number
More than 10 numbers, contact manufacturer.
Model or catalog number
- 9031
- 9051
Companies
- Manufacturer
-
Merck Millipore Ltd.
Tullagreen
Carrigtwohill, Cork
T45 KD29
IRELAND