This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Microscan Rapid Neg Urine Combo Panel (2014-10-22)

Starting date:
October 22, 2014
Posting date:
November 7, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-42077

Recalled Products

Microscan Rapid Neg Urine Combo Panel

Reason

There is an increase in false susceptible results for Enterobacteriaceae, Acinetobacter species and Pseudomonas aeruginosa isolates on Rapid Neg BP combo Type 3 (B1017-117)and Rapid Neg Urine combo Type 1 (B1017-167) panels, as compared to frozen reference panels, with the four key drugs evaluated (Aztreonam, Cefotaxime, Ceftazidime and Ceftriaxone). The false susceptible results could lead to the potential treatment of patients with an ineffective antibiotic.

Affected products

Microscan Rapid Neg Urine Combo Panel

Lot or serial number

2014-11-01

2015-01-03

2015-02-14

2015-03-21

2015-05-20

2015-06-18

2015-07-18

2015-08-11

Model or catalog number

B1017-167

Companies
Manufacturer

Siemens Healthcare Diagnostics Inc.

2040 Enterprise Blvd. West

Sacramento

95691

California

UNITED STATES