MicroScan Pos Combo Panels 29, 33, and 34, MicroScan Pos BP Combo Panel Type 20 (2019-03-20)
- Starting date:
- March 20, 2019
- Posting date:
- April 12, 2019
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-69538
Last updated: 2019-04-12
Affected Products
- MicroScan Pos Combo Panels 29, 33, and 34
- MicroScan Pos BP Combo Panel Type 20
Reason
Beckman Coulter has confirmed customer reports of insufficient growth observed in growth well in overnight conventional gram-positive panel lots with quality control (QC) organism M. Luteus ATCC 49732. The expected result is growth.
The issue was traced to a subassembly used in the manufacturing of the affected panel lots.
Affected products
-
MicroScan Pos Combo Panels 29, 33, and 34
Lot or serial number
43749
43753
43754
43756
Model or catalog number
B1017-209
B1017-211
B1017-214
Companies
- Manufacturer
-
Beckman Coulter, Inc.
250 S. Kraemer Blvd.
Brea
92821
California
UNITED STATES
B. MicroScan Pos BP Combo Panel Type 20
Lot or serial number
43754
Model or catalog number
B1017-202
Companies
- Manufacturer
-
Beckman Coulter, Inc.
250 S. Kraemer Blvd.
Brea
92821
California
UNITED STATES