Health product recall

Microscan Pos Combo Panel Type 43 and Mic Panel Type 34 (2017-04-17)

Starting date:
April 17, 2017
Posting date:
May 16, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-63294

Affected Products

A. Microscan Pos Combo Panel Type 43
B. Microscan Pos Mic Panel Type 34

Reason

Beckman coulter did not inform the customer that TigeCycline minimum inhibitory concentration (MIC) is being transmitted to the LIS interface data stream. In order to support communication with all LIS systems, the LabPro software is transmitting the TigeCycline MIC value. Some LIS systems can only accept numerical values for antimicrobial agent test results and any interpretive criteria applied to these numerical values are generated by the LIS system.

Affected products

A. Microscan Pos Combo Panel Type 43

Lot or serial number

Not Lot Specific

Model or catalog number

B1017-217

Companies
Manufacturer
Beckman Coulter Inc.
250 S. Kraemer Blvd.
Brea
92821
UNITED STATES

B. Microscan Pos Mic Panel Type 34

Lot or serial number

Not Lot Specific

Model or catalog number

B1017-216

Companies
Manufacturer
Beckman Coulter Inc.
250 S. Kraemer Blvd.
Brea
92821
UNITED STATES