This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Microplate Reader Infinite (2015-06-15)

Starting date:
June 15, 2015
Posting date:
June 15, 2015
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-53619

Recalled Products

  1. Microplate Reader Infinite 200 PRO
  2. Microplate Reader Infinite M 1000 PRO
  3. Microplate Reader Infinite F500
  4. Microplate Reader Infinite M1000
  5. Microplate Reader Infinite 200

Reason

During an internal project Tecan has determined that due to a firmware bug incorrect luminescence measurement values are displayed if the actual luminescence measurement value equals 4096 RLU or multiples thereof. It was determined that these incorrect measurement values may potentially impact the application result.

Affected products

A. Microplate Reader Infinite 200 PRO

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number
  • 30050303
Companies
Manufacturer

Tecan Austria GmbH,

Untersbergstrasse 1A, A-5082,

Grodig

9999

AUSTRIA



B. Microplate Reader Infinite M 1000 PRO

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number
  • 30063849
Companies
Manufacturer

Tecan Austria GmbH,

Untersbergstrasse 1A, A-5082,

Grodig

9999

AUSTRIA



C. Microplate Reader Infinite F500

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number
  • 30019337
Companies
Manufacturer

Tecan Austria GmbH,

Untersbergstrasse 1A, A-5082,

Grodig

9999

AUSTRIA



D. Microplate Reader Infinite M1000

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number
  • 30034301
  • 30061442
Companies
Manufacturer

Tecan Austria GmbH,

Untersbergstrasse 1A, A-5082,

Grodig

9999

AUSTRIA



E. Microplate Reader Infinite 200

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number
  • 30016056
Companies
Manufacturer

Tecan Austria GmbH,

Untersbergstrasse 1A, A-5082,

Grodig

9999

AUSTRIA