Health product recall

Micro SSP High Resolution HLA Class II DNA Typing Trays (2013-11-04)

Starting date:
November 4, 2013
Posting date:
February 3, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-37535

Recalled Products

Micro SSP High Resolution HLA Class II DNA Typing Trays 

Reason

The use of this product in combination with optional analysis software, HLA Fusion software, may cause misassignment to DRB1*15:65 when positive reactivity is seen in well 11A.  With the presence of DRB4*01:03 and DRB5*01:01, only a DRB1*15:65 without DRB1*15:01 typing result is suggested.  This is due to a limitation in the software with cross loci wells.  Analysis of this product without HLA fusion software will not generate incorrect suggested typing results.

Affected products

Micro SSP High Resolution HLA Class II DNA Typing Trays

Lot or serial number

Lot 005, batches 003-034

Model or catalog number

SSP2H

Companies
Manufacturer

One Lambda Inc.

21001 Kittridge Street

Canoga Park

91303-2801

California

UNITED STATES