This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Micro SSP DNA Typing Trays (2013-02-21)

Starting date:
February 21, 2013
Posting date:
February 10, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-37841

Recalled Products

Micro SSP DNA Typing Trays

Reason

Null allele information on the lot specific documents of these products is incorrect and may cause mis-assignment - refer to table 1- reaction patterns and potential mis-assignments.  The root cause investigations determined that null allele entries during nomenclature update were not verified per standard operating procedure.  No adverse events or product complaints have been received.

Affected products

Micro SSP DNA Typing Trays

Lot or serial number

SSP1-11, lot 002,

SSP1-26, lot 01A,

SSP2H, lot 005,

SSP2H, lot 04A,

SSP2L, lot 007,

SSP2LB, lot 004,

SSPABDR, lot 008,

SSPABDR, lot 009,

SSPABDRX, lot 008,

SSPABDRX, lot 009,

SSPDRQP1, lot 001,

SSPML02, lot 003,

SSPML02, lot 004,

SSPR1-11, lot 003

Model or catalog number

SSP1-11

SSP1-26

SSP2H

SSP2L

SSP2LB

SSPABDR

SSPABDRX

SSPDRQP1

SSPML02

SSPR1-11

Companies
Manufacturer

One Lambda Inc.

21001 Kittridge Street

Canoga Park

91303-2801

California

UNITED STATES