Merlin Patient Care System - Programmer Software
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products | Lot or serial number | Model or catalogue number |
---|---|---|
Merlin Patient Care System - Programmer Software | version v25.3.x or earlier | 3330 VER 25.2.4 3330 VER 24.4.4 3330 VER 18.2 3330 VER 24.6.4 3330 VER 24.3.4 3330 VER 17.2.1 3330 VER 23.0.4 3330 VER 25.0.4 3330 VER 25.3.4 3330 VER 24.2.4 3330 VER 20.0.1.2 3330 VER 24.5.4 |
Issue
Abbott is notifying customers of the potential for programmer software (Merlin pcs) and remote monitoring software applications (Merlin.net) to display overestimated predicted battery longevity for certain pacemakers. Pacemaker/battery functionality, therapy delivery, and longevity remain normal and within specifications. Voltage measurements and ERI (elective replacement indicator), which is based on direct voltage measurement, remain accurate.
Recall start date: June 16, 2022