Merlin Patient Care System - Programmer Software
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalogue number |
---|---|---|
Merlin Patient Care System - Programmer Software |
version v25.3.x or earlier |
3330 VER 25.2.4 |
Issue
Abbott is notifying customers of the potential for programmer software (Merlin pcs) and remote monitoring software applications (Merlin.net) to display overestimated predicted battery longevity for certain pacemakers. Pacemaker/battery functionality, therapy delivery, and longevity remain normal and within specifications. Voltage measurements and ERI (elective replacement indicator), which is based on direct voltage measurement, remain accurate.
Recall start date: June 16, 2022
Additional information
Details
St. Jude Medical, Cardiac Rhythm Management Division