Health product recall

Merit basixCOMPAK™, PhD™ Hemostasis Valve, and basixSKY™ Inflation Syringe

Last updated

Summary

Product
Merit basixCOMPAK™, PhD™ Hemostasis Valve, and basixSKY™ Inflation Syringe
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalog number

basixCOMPAK™ MAP802

I3161210

IN4802

basixCOMPAK™ MAP802

I3161211

IN4802

PhD™ Hemostasis Valve

I3146099

MAP801

PhD™ Hemostasis Valve

I3186946

MAP801

PhD™ Hemostasis Valve

I3219783

MAP802

PhD™ Hemostasis Valve

I3209935

MAP802

PhD™ Hemostasis Valve

I3215926

MAP802

PhD™ Hemostasis Valve

I3192937

MAP802

PhD™ Hemostasis Valve

I3215925

MAP802

PhD™ Hemostasis Valve

I3129726

MAP802

PhD™ Hemostasis Valve

I3129725

MAP802

PhD™ Hemostasis Valve

I3219784

MAP800

PhD™ Hemostasis Valve

I3206933

MAP800

PhD™ Hemostasis Valve

I3219792

MAP800

PhD™ Hemostasis Valve

I3186951

MAP800

PhD™ Hemostasis Valve

I3191334

MAP800

Custom Procedure Kit - Stroke Kit

T3163727

K12T-10798A

Custom Procedure Kit - Stroke Kit

T3217266

K12T-10798A

basixSKY™ Inflation Syringe MAP802

H3155575

SKY1802

basixSKY™ Inflation Syringe MAP802

H3141576

SKY1802

basixSKY™ Inflation Syringe MAP802

H3155576

SKY1802

basixSKY™ Inflation Syringe MAP802

H3155577

SKY1802

basixSKY™ Inflation Syringe MAP802

H3079946

SKY1802

basixSKY™ Inflation Syringe MAP802

H3141577

SKY1802

Issue

Merit Medical became aware of a manufacturing defect after an increased number of complaints. Investigation into the issue concluded that the lack of fluorinated silicone in the bleed back seal (the secondary valve) caused by the mix of incorrect silicone during production line influenced the performance of the PhD™ Hemostasis Valve. The lack of silicone was found to prevent the secondary valve from releasing and closing properly. The valve cap potentially would remain depressed causing leakage.

Recall start date: July 30, 2025

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Cardiovascular
Health products - Medical devices - General hospital and personal use
Companies

Merit Medical Systems Inc.

1600 West Merit Parkway, South Jordan, Utah, United States, 84095

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-77904