Health product recall

Merge PACS (2020-08-03)



Starting date:
August 3, 2020
Posting date:
August 28, 2020
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-73781


Last updated:

Affected Products

Merge PACS

Reason

  • Issue 1: pixel value measurements from the probe, ROI, and region analysis area tools are incorrect on multi-planar reformats (mprs) generated by Merge PACS.
  • Issue 2: black dots may be displayed in a scattered pattern throughout images.
  • Issue 3: software correction to address a incorrect part number reference on the about screen of software version 8.1.2 (References part number for v8.1.1).

Affected products

Merge PACS

Lot or serial number
  • v.8.1
  • v.8.1.1
  • v.8.1.2
Model or catalog number

MERGE PACS

Companies
Manufacturer

Merge Healthcare Incorporated,

900 Walnut Ridge Drive,

Hartland

53029

Wisconsin

UNITED STATES