This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Mentor Gel Sizers (2014-05-27)

Starting date:
May 27, 2014
Posting date:
June 12, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-40001

Recalled products

  1. Mentor Gel Sizers 

Reason

A complaint was received for a volume discrepancy between the pad print (500cc) and laser engraved marking and all other product labeling (550cc) on a Mentor Gel Sizer. During the investigation, a second lot was identified as having a style discrepancy between the pad print moderate profile(MP) and all other product labeling moderate plus(M+).

Affected products

A. Mentor Gel Sizers (2014-05-27)

Lot or serial number
  • 6680501
  • 6749298
Model or catalog number
  • RSZ-2751
  • RSZ-5501
Companies
Manufacturer

Mentor Medical Systems

B.V., Zernikedreef 2

Leiden, Netherlands

2333 CL