Health product recall

Meniscal Knot Pusher and Suture Cutter (2015-05-08)

Starting date:
May 8, 2015
Posting date:
June 3, 2015
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-53643

Recalled Products

A. Meniscal Knot Pusher and Suture Cutter

Reason

A Smith & Nephew internal investigation found the packaging material and design to be insufficient in preventing the Ultra FAST FIX knot pusher/cutter's sharp edges from puncturing the packaging pouch and rendering the device as non-sterile.

Affected products

A. Meniscal Knot Pusher and Suture Cutter

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

72201537

Companies
Manufacturer

Smith & Nephew Inc.

Endoscopy Division

150 Minuteman Road

Andover, Massachusetts

01810

UNITED STATES