MEDRAD® Centargo Injection System
Brand(s)
Summary
Customers will be notified with instructions to follow.
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
---|---|---|
MEDRAD® Centargo Injection System |
Contact the manufacturer. |
CENT-SYS-BAT |
Issue
Centargo patient line sensor is a component of the Centargo injection system which detects a patient line. An investigation into reported air injection concerns found that, under certain special conditions, the patient line sensor may occasionally have difficulty detecting the presence of the patient line. This intermittent detection issue, combined with user reliance on automated system functions (such as auto-prime) and possible workflow inconsistencies, may slightly increase the likelihood of small air injections (1-10 ml). However, the overall risk remains low.
Recall start date: December 13, 2024
Additional information
Details
Imaxeon Pty Ltd
Unit 1, 38-46 South Street, Rydalmere, New South Wales, Australia, 2116
Get notified
Receive notifications for new and updated recalls and alerts by category.