Health product recall

Medrad Intego Pet Infusion System - Source Admini (2015-11-30)

Starting date:
November 30, 2015
Posting date:
February 1, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-56884

Affected Products

MEDRAD INTEGO PET INFUSION SYSTEM - SOURCE ADMINI

Reason

The inner diameter of the saline tubing is out of specification and may result in the system failing to position the dose correctly during extraction, potentially leading to a low measurement of the extracted dose or to a higher dose being extracted. The occluded SAS tubing may impair the priming function and prompt an "RP prime failed. Check SAS and vial installation, then reprime" error message.

Affected products

MEDRAD INTEGO PET INFUSION SYSTEM - SOURCE ADMINI

Lot or serial number

More than 10 lots, contact manufacturer

Model or catalog number

INT CSS

Companies
Manufacturer
Bayer Medical Care Inc.
1 Bayer Drive
Indianola
15051-0780
Pennsylvania
UNITED STATES