MEDRAD INTEGO PET INFUSION SYSTEM - SOURCE ADMINI
- Starting date:
- October 5, 2016
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Identification number:
- RA-60678
Affected Products
Reason
On November 2015, during a review of historical Intego SAS Testing, it was discovered that documented test evidence related to 85 PSI (Stability Index) for 35 seconds could not be found. When the necessary testing was recently conducted, some failures were found at three year simulated age testing. The failed samples exhibited leaks in the coil bond joint. There was no test failures of the samples at one year accelerated age. Product is currently labeled with a 3 year shelf-life and thus needs to be corrected until this shelf-life can be re- established.
Affected products
MEDRAD INTEGO PET INFUSION SYSTEM - SOURCE ADMINI
Lot or serial number
More than 10 numbers, contact manufacturer.
Model or catalog number
INT CSS
Companies
- Manufacturer
-
Bayer Medical Care Inc.
1 Bayer Drive
Indianola
15051-0780
Pennsylvania
UNITED STATES