This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

MEDRAD INTEGO PET INFUSION SYSTEM - SOURCE ADMINI

Starting date:
October 5, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Identification number:
RA-60678

Affected Products

Reason

On November 2015, during a review of historical Intego SAS Testing, it was discovered that documented test evidence related to 85 PSI (Stability Index) for 35 seconds could not be found. When the necessary testing was recently conducted, some failures were found at three year simulated age testing. The failed samples exhibited leaks in the coil bond joint. There was no test failures of the samples at one year accelerated age. Product is currently labeled with a 3 year shelf-life and thus needs to be corrected until this shelf-life can be re- established.

Affected products

MEDRAD INTEGO PET INFUSION SYSTEM - SOURCE ADMINI

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

INT CSS

Companies
Manufacturer

Bayer Medical Care Inc.

1 Bayer Drive

Indianola

15051-0780

Pennsylvania

UNITED STATES