Health product recall

McGRATH™ Mac 2 Laryngoscope Blade

Brand(s)
Last updated

Summary

Product
McGRATH™ Mac 2 Laryngoscope Blade
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalog number

McGRATH™ Mac 2 Laryngoscope Blade

22102001
22082207
22083101

350-017-000
350-084-000

Issue

Medtronic received six (6) complaints of poor picture quality and foggy and/or blurred images on the McGRATH™ Mac 2 video laryngoscope screen during use of these blades. The reports indicated a delay to treatment while an alternate device was located and/or oral tissue injury due to multiple intubation attempts. Medtronic investigation of these customer reports identified a manufacturing issue which resulted in an ineffective application of the antifog agent on the blades. An ineffective application of the anti-fog coating can result in condensation accumulating on the blade during use, resulting in the image on the video laryngoscope screen being blurred or obscured.

Recall start date: August 8, 2023

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Anaesthesiology
Companies

Covidien LLC

15 Hampshire Street, Mansfield, Massachusetts, United States, 02048

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-74924

Get notified

Receive emails about new and updated recall and safety alerts.

Subscribe