Issue

Manufacturer initiated recall notifying users of variability of only the level 1 troponin I values included in the Mas CardioImmune XL controls (confirmed after 8 complaints). The level 1 control is included in lots of control kits CAI-XL 1, CAI-XL4  and CAI-XLS. This variability may lead some vials to recover higher than expected values, therefore manufacturer directs user to not use troponin I values for level 1 (other constituents in level 1 are not affected).

Recall start date: January 5, 2023

Additional information

Details
Original published date: 2023-01-24
Alert / recall type
Health product recall
Category
Health products - Medical devices - Chemistry
Companies

Microgenics Corporation

46500 Kato Road, Fremont, California, United States, 94538

Published by
Health Canada
Audience
General public
Healthcare
Industry
Recall class
Type II
Identification number
RA-72612