Health product recall

Maquet Surgical Lights

Brand(s)
Last updated

Summary

Product
Maquet Surgical Lights
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalog number

Volista Surgical Light

All lots.

ARDXXXXXXXXXX

Lucea 10/40

All lots.

ARDXXXXXXXXXX

Lucea 50/100

All lots.

ARDXXXXXXXXXX

PrismAlix

All lots.

ARDXXXXXXXXXX

Maquet Equipment

All lots.

ARDXXXXXXXXXX

Axcel+

All lots.

ARDXXXXXXXXXX

Hanaulux HLX3000

All lots.

ARDXXXXXXXXXX

Blueline Series 30/80

All lots.

ARDXXXXXXXXXX

Power LED 300 Surgical Light

All lots.

ARDXXXXXXXXXX

Axcel

All lots.

ARDXXXXXXXXXX

H Led Surgical Light

All lots.

ARDXXXXXXXXXX

Blue Series 30/80 Surgical Light

All lots.

ARDXXXXXXXXXX

Maquet Power LED II

All lots.

ARDXXXXXXXXXX

Power LED Surgical Light

All lots.

ARDXXXXXXXXXX

X'Ten Surgical Light

All lots.

ARDXXXXXXXXXX

Issue

Maquet S.A.S. /Getinge has received a customer complaint regarding a fall of a light system in an operating room in Hong Kong (no injuries reported). Tnvestigation into this event demonstrated that the recommendation of maintenance and service were not performed as instructed in the current user and maintenance manuals. This notification communicates retrospectively current maintenance instructions and programs to customers.

Recall start date: November 29, 2023

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General and plastic surgery
Companies

Maquet S.A.

Avenue De La Pomme De Pin, Orleans, France, 45074

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-74719

Get notified

Receive emails about new and updated recall and safety alerts.

Subscribe