Magnetom Systems
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
---|---|---|
Magnetom Lumina |
196408 |
11344916 |
Magnetom Skyra |
45186 |
10432915 |
Magnetom Sola Fit |
202567 |
11410482 |
Magnetom Vida |
More than 10 numbers, contact manufacturer. |
11060815 |
Magnetom Altea |
189602 |
11344915 |
Magnetom Sola |
More than 10 numbers, contact manufacturer. |
11291455 |
Magnetom Aera |
More than 10 numbers, contact manufacturer. |
10432914 |
Magnetom Prisma |
166026 |
10849582 |
Magnetom Avanto Fit |
211019 |
11516216 |
Issue
Sporadically, the distance measurement values displayed in MRI images on third-party postprocessing systems might be wrong if:
- enhanced MRI images with rectangular FOV (field of view) are created at the MR scanner, and
- in Syngo.via or MR view&go, the user rotates the images to display them in any orientation and saves them, and
- afterwards, the user exports the images to another Dicom node using MR export rules (interoperability mode) the distance measurement values on newly created series might be displayed incorrectly on systems that receive these new images.
Recall start date: December 21, 2022
Additional information
Details
Siemens Healthcare Gmbh
Henkestr. 127, Erlangen, Germany, 91052
Get notified
Receive notifications for new and updated recalls and alerts by category.