Health product recall

Magnetom Scanner

Last updated

Summary

Product
Magnetom Scanner
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalogue number

Magnetom Scanner

More than 10 numbers, contact manufacturer.

11060815

Magnetom Scanner

196408

11344916

Magnetom Scanner

196427

11344916

Magnetom Scanner

196822

11344916

Magnetom Scanner

196406

11344916

Magnetom Scanner

196680

11344916

Issue

The service pack NX VA60A-SP04 is a bug fix update (full installation) that solves/improves the following:

  • Latest it security deliveries and increase in overall stability (syngo MR XA60A-SD12 content included)
  • Installation of multiple SW delta packages now possible at the same time
  • EU DataAact (EY) 2023/2854 implementation:access to all machine data
  • Solved image loading errors for MR Breast Biopsy
  • General MR Protocol Manager improvements
  • Coil specifications adjusted for BM Body 18 Pilot Tone and Spine 32 and 1.5T Sentinelle Breast coils image brightness
  • Supporting IEC 60601-1 compliance, enabling predictable and safe system behavior with regard to table movement

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Radiology
Companies
Siemens Healthineers Ag
Siemensstr. 3, Forchheim, Germany, 91301
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Recall date
Identification number
RA-82871

Get notified

Receive emails about new and updated recall and safety alerts.

Subscribe