Magnetom MRI System
Brand(s)
Last updated
Summary
Product
Magnetom MRI System
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Affected products
Affected Products | Lot or serial number | Model or catalog number | ||
---|---|---|---|---|
Magnetom Lumina | 196408 196427 196406 | 11344916 | ||
Magnetom Vida Fit | 202080 | 11410481 | ||
Magnetom Vida | More than 10 numbers, contact manufacturer. | 11060815 | ||
Magnetom Prisma Fit | 167017 167006 167068 67090 167104 167096 167057 67007 167084 | 10849583 | ||
Tim 4G - Avantofit | 168013 168009 | 10849578 10849579 | ||
Magnetom Terra | 79117 | 10882764 | ||
Magnetom Sola | More than 10 numbers, contact manufacturer. | 11291455 | ||
Tim 4G - Skyra Fit | 70007 170089 70006 170006 | 10849580 | ||
Magnetom Altea | 189602 189915 189956 189700 190027 190546 | 11344915 | ||
Magnetom Aera | More than 10 numbers, contact manufacturer. | 10432914 | ||
Magnetom Prisma | 166026 166139 166247 66060 66071 166208 166092 66049 66108 | 10849582 | ||
Magnetom Avanto Fit | 211077 211050 211067 211019 | 11516216 | ||
Magnetom Skyra | More than 10 numbers, contact manufacturer. | 10432915 | ||
Magnetom Sola Fit | 202567 202589 202562 202570 202619 202609 202617 | 11410482 |
Issue
Siemens Healthcare will be informing customers about potential dot- or line-shaped artifacts caused by the presence of lubricating grease within the headband of the over-ear headphones listed below when used during head examinations with the Magnetom MRI systems.
Recall Start Date: August 28, 2023
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Radiology
Companies
Siemens Healthcare Gmbh
Henkestr. 127, Erlangen, Germany, 91052
Published by
Health Canada
Audience
General public
Healthcare
Industry
Recall class
Type II
Identification number
RA-74246