This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

MacroLyte Premie Dispersive Electrode(2014-05-12)

Starting date:
May 12, 2014
Posting date:
June 5, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-39813

Recalled Products

MacroLyte Premie Dispersive Electrode

Reason

Through the Conmed corporation complaint handling system, Conmed received six (6) complaints regarding sparking, no output and burning at the cord set connection. The initial complaint was received on 04/11/2014. The complaints were for the gel possibly making contact with an internal staple and this may cause no output, sparking and heat buildup between the staple and foil contact surface. One patient in Japan was reportedly to have received a burn. Conmed confirmed instances where the gel contact with the internal staple causes the staple to corrode over time. Corroded staples may fail upon device activation, causing a burn through the foam which covers the staple.

Affected products

MacroLyte Premie Dispersive Electrode

Lot or serial number

More than 10 numbers, contact manufacturer. 

Model or catalog number

440-2400

 

Companies
Manufacturer

Conmed Corporation

525 French Road

Utica

13502

New York

UNITED STATES