M Zn-Air Battery Pak
Brand(s)
Summary
Follow manufacturer's instructions on removing non-tamperproof battery door and dispose of it. Ask for replacements with tamperproof battery doors.
Affected products
Affected products |
Lot or serial number |
Model or catalogue number |
|---|---|---|
M Zn-Air Battery Pak |
All lots. |
CI-5501-150 |
M Zn-Air Battery Pak |
All lots. |
CI-5501-260 |
M Zn-Air Battery Pak |
All lots. |
CI-5501-280 |
M Zn-Air Battery Pak |
All lots. |
CI-5501-110 |
M Zn-Air Battery Pak |
All lots. |
CI-5501-190 |
M Zn-Air Battery Pak |
All lots. |
CI-5501-250 |
M Zn-Air Battery Pak |
All lots. |
CI-5501-120 |
M Zn-Air Battery Pak |
All lots. |
CI-5501-240 |
M Zn-Air Battery Pak |
All lots. |
CI-5501-270 |
M Zn-Air Battery Pak |
All lots. |
CI-5501-130 |
M Zn-Air Battery Pak |
All lots. |
CI-5501-140 |
Issue
The non‑tamperproof battery door in the M Zn‑Air Battery Pak (CI‑5501) does not meet IEC 60601‑1‑11:2015 clause 8.5.2 because it allows battery access without a tool, creating potential safety risk. No injuries are reported, but the recall instructs customers to dispose of the affected door per local guidelines.
Additional information
Details
Advanced Bionics, LLC
28515 Westinghouse Place, Valencia, California, United States, 91355
Get notified
Receive emails about new and updated recall and safety alerts.