Lyra SARS-CoV-2 Assay (M120) (2021-04-27)
- Starting date:
- April 27, 2021
- Posting date:
- July 20, 2021
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type I
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-76061
Last updated:
2021-07-20
Affected Products
Lyra SARS-CoV-2 Assay (M120)
Reason
Quidel has identified that rare situations patients infected with high viral load of SARSâCoVâ2 may have samples that generate Ct values lessâthan or equal to (â¤) 5 when certain amplification and detection platforms are used. Based on the interpretation table in the package insert these specimens would be interpreted as negative.
Affected products
Lyra SARS-CoV-2 Assay (M120)
Lot or serial number
All lots.
Model or catalog number
M120
Companies
- Manufacturer
-
Diagnostic Hybrids, Inc.
2005 EAST STATE STREET, SUITE 100
45701