Health product recall

Lyra SARS-CoV-2 Assay (M120) (2021-04-27)



Starting date:
April 27, 2021
Posting date:
July 20, 2021
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type I
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-76061


Last updated:

Affected Products 

Lyra SARS-CoV-2 Assay (M120)

Reason

Quidel has identified that rare situations patients infected with high viral load of SARS‐CoV‐2 may have samples that generate Ct values less‐than or equal to (≤) 5 when certain amplification and detection platforms are used. Based on the interpretation table in the package insert these specimens would be interpreted as negative.

Affected products

Lyra SARS-CoV-2 Assay (M120)

Lot or serial number

All lots. 

Model or catalog number

M120

Companies
Manufacturer

Diagnostic Hybrids, Inc.

2005 EAST STATE STREET, SUITE 100

45701