This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Linear X-Ray Collimator and Progeny Vantage Panoramic X-Ray System (2014-11-19)

Starting date:
November 19, 2014
Posting date:
April 27, 2015
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-53023

Recalled Products

A. Linear MC150 Collimator

B. Linear II Collimator

C. Progeny Vantage Panoramic X-ray System with Cephalometric Extensions with Sensor

D. Progeny Vantage Panoramic X-ray System Cephalometric Extensions with Sensor

 

Reason

Device label failed to pass the JEC 60601-1 durability test and the affected device label may lose legibility during the useful life of the device.

Affected products

A. Linear MC150 Collimator

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

70-63XXX

Companies
Manufacturer

Midmark Corporation

675 Heathrow Drive

Lincolnshire

60069

Illinois

UNITED STATES



B. Linear II Collimator

Lot or serial number

FA26112

Model or catalog number

70-43XXX

Companies
Manufacturer

Midmark Corporation

675 Heathrow Drive

Lincolnshire

60069

Illinois

UNITED STATES



C. Progeny Vantage Panoramic X-ray System with Cephalometric Extensions with Sensor

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

V5000

Companies
Manufacturer

Midmark Corporation

675 Heathrow Drive

Lincolnshire

60069

Illinois

UNITED STATES



D. Progeny Vantage Panoramic X-ray System Cephalometric Extensions with Sensor

Lot or serial number

1Z08011

1Z08012

IZ08013

Model or catalog number

C6000

Companies
Manufacturer

Midmark Corporation

675 Heathrow Drive

Lincolnshire

60069

Illinois

UNITED STATES