This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Lifecodes HLA-DRB1 Typing Kit (2014-01-29)

Starting date:
January 29, 2014
Posting date:
March 18, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-38317

Recalled Products

Lifecodes HLA-DRB1 Typing Kit 

Reason

Immucor Transplant Diagnostics, Inc. has confirmed HLA-DRB1 Probe 091 has been designed with the incorrect sequence. As a result, the Probe is not expected to react with any alleles, including the three rare alleles for which it was designed to detect: DRB107:20 and 11:86 and 13:16. Analysis with Probe 091 as a false negative does not cause a change in serological typing or level of resolution.

Affected products

Lifecodes HLA-DRB1 Typing Kit 

Lot or serial number

05073D

08021A

08021AIVD

10102X

Model or catalog number

628751-50

Companies
Manufacturer

Immucor Transplant Diagnostics, Inc.

550 West Avenue

Stamford

06902

Connecticut

UNITED STATES