Health product recall

LIFECODES HLA-DRB1 SSO, HLA-DQA1/B1 SSO, and HLA-DPA/B1 SSO Typing Kit (2017-07-07)

Starting date:
July 7, 2017
Posting date:
March 23, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-66286

Affected products

  1. LIFECODES HLA-DRB1 SSO TYPING KIT
  2. LIFECODES HLA-DQA1/B1 SSO TYPING KIT
  3. LIFECODES HLA-DPA1/B1 SSO TYPING KIT

Reason

As a result of a complaint investigation, Immucor found that a number of its LIFECODES HLA SSO TYPING KITS were affected by an error in the MATCH IT! DNA 3.27.0 allele database update. The lot specific software files for these kits had incorrect probe assignments for some alleles which could have led to a potential mistype of certain alleles.

Affected products

A. LIFECODES HLA-DRB1 SSO TYPING KIT

Lot or serial number
  • 04226A
  • 11125A
Model or catalog number

628923

Companies
Manufacturer

Immucor GTI Diagnostics, Inc.,

20925 Crossroads Circle,

Waukesha

53186

UNITED STATES

B. LIFECODES HLA-DQA1/B1 SSO TYPING KIT

Lot or serial number
  • 06306A
  • 10205A
  • 3004970
Model or catalog number

628930

Companies
Manufacturer

Immucor GTI Diagnostics, Inc.,

20925 Crossroads Circle,

Waukesha

53186

UNITED STATES

C. LIFECODES HLA-DPA1/B1 SSO TYPING KIT

Lot or serial number
  • 12085A
  • 3004756
Model or catalog number

628936

Companies
Manufacturer

Immucor GTI Diagnostics, Inc.,

20925 Crossroads Circle,

Waukesha

53186

UNITED STATES