Health product recall

LIFECODES HLA-DQA1/B1 Typing Kit (2015-12-22)

Starting date:
December 22, 2015
Posting date:
January 14, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-56644

Affected Products

A. LIFECODES HLA-DQA1/B1 Typing Kit

Reason

Immucor Transplant Diagnostics Inc., is investigating a single instance of false positive reactivity of probe dqb253 associated with an allele pair of DQB1*03:03:02:01, DQB1*03:04:01.  A false positive result causes the exclusion of DQB1*03:03:02:01/ 03:04:01 as a possible allele combination.  The false positive reactivity is only observed in a DQB1*03:03:02:01, DQB1*03:04:01 heterozygous combination. Other combinations with DQB1*03:03:02:01 or DQB1*03:04:01 are not affected.

Affected products

A. LIFECODES HLA-DQA1/B1 Typing Kit

Lot or serial number
  • 07065B
  • 11184C
Model or catalog number

628930

Companies
Manufacturer
Immucor Transplant Diagnostics, Inc.
550 West Avenue
Stamford
06902
Connecticut
UNITED STATES