This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

LIFECODES HLA-DQA1/B1 Typing Kit (2015-12-22)

Starting date:
December 22, 2015
Posting date:
January 14, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-56644

Affected Products

A. LIFECODES HLA-DQA1/B1 Typing Kit

Reason

Immucor Transplant Diagnostics Inc., is investigating a single instance of false positive reactivity of probe dqb253 associated with an allele pair of DQB1*03:03:02:01, DQB1*03:04:01.  A false positive result causes the exclusion of DQB1*03:03:02:01/ 03:04:01 as a possible allele combination.  The false positive reactivity is only observed in a DQB1*03:03:02:01, DQB1*03:04:01 heterozygous combination. Other combinations with DQB1*03:03:02:01 or DQB1*03:04:01 are not affected.

Affected products

A. LIFECODES HLA-DQA1/B1 Typing Kit

Lot or serial number
  • 07065B
  • 11184C
Model or catalog number

628930

Companies
Manufacturer

Immucor Transplant Diagnostics, Inc.

550 West Avenue

Stamford

06902

Connecticut

UNITED STATES