This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Lifecodes HLA-C eRES SSO Typing Kit (2016-05-02)

Starting date:
May 2, 2016
Posting date:
May 16, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-58394

Affected products 

  1. Lifecodes HLA-C eRES SSO Typing Kit

Reason

It has been determined that the threshold set for HLA-C eRES probes 380, 382, 384 and 385 are set high and may result in false negative probe results. The probes resolve the following CWD alleles probe 380 resolves c*03:06, probe 382 resolves c*12:13, probe 384 resolves c*01:08 and probe 385 resolves c*08:06.

Affected products

A. Lifecodes HLA-C eRES SSO Typing Kit

Lot or serial number

Lot # 10235B

Model or catalog number
  • 628921
Companies
Manufacturer
Immucor Transplant Diagnostics, Inc.
550 West Avenue
Stamford
06902
Connecticut
UNITED STATES