This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Liaison EBV IgM (2015-08-04)

Starting date:
August 4, 2015
Posting date:
October 22, 2015
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-55474

Affected Products

Liaison EBV IgM

Reason

Recall initiated due to observed trend which may cause the potential for false positive results. Internal testing confirmed the kit performance is meeting performance claims. EBV-IgM (Esptein-Barr Virus Immunoglobulin) results should not be used alone but in conjunction with other EBV serological markers to determine disease state.

Affected products

Liaison EBV IgM

Lot or serial number

028036X

Model or catalog number

310500

Companies
Manufacturer

Diasorin Inc.

1951 Northwestern Avenue, P.O. Box 285

Stillwater

55082

Minnesota

UNITED STATES