This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Liaison Aldosterone control sets (2014-12-05)

Starting date:
December 5, 2014
Posting date:
December 18, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-42935

Recalled Products

A) Liaison Aldosterone control set (588116)

B) Liaison Aldosterone control set (588120)

Reason

The Liaison Aldosterone Control Set IFU contains an error in storage and stability.  The control open use stability claim is 6 weeks when stored at 2-8c but the correct control open use stability is 4 weeks when properly stored at 2-8c.

Affected products

A) Liaison Aldosterone control set (588116)

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

310451

Companies
Manufacturer

Diasorin Inc.

1951 Northwestern Ave. P.O. Box 285

Stillwater

55082

Minnesota

UNITED STATES



B) Liaison Aldosterone control set (588120)

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

310451

Companies
Manufacturer

Diasorin Inc.

1951 Northwestern Ave. P.O. Box 285

Stillwater

55082

Minnesota

UNITED STATES