Health product recall

Levo Arm - Component of the Levo Head Positioning System (2019-10-21)

Starting date:
October 21, 2019
Posting date:
November 8, 2019
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-71551



Last updated: 2019-11-08

Affected Products

Levo Arm - Component of the Levo Head Positioning System

Reason

Each of these packages includes a Levo Arm (7887-050) that provides the primary positioning and locking capabilities of the system. Recently, we became aware of the potential for fluid ingress resulting from unapproved cleaning techniques. Examples of incorrect cleaning could include submersion or exposure of the unit to direct fluid flow. This may result in fluid entering the unit, rendering the internal electronics inoperable and leading to a failure of the arm's locking mechanism in isolated situations.

Affected products

Levo Arm - Component of the Levo Head Positioning System

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

7887-050

Companies
Manufacturer

Mizuho Orthopedic Systems, Inc.

30057 Ahern Avenue

Union City

94587-1234

California

UNITED STATES