Leksell Gamma Knife Perfexion (2014-10-06)
- Starting date:
- October 6, 2014
- Posting date:
- October 10, 2014
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-41739
Recalled Products
Leksell Gamma Knife Perfexion
Reason
We have got a report that the latches may be locked before they have been fully turned, resulting in a poorly locked frame. Investigations show that it may be possible to have a situation where the adapter latches can get stuck just above the coordinate frame surface and as a result the coordinate frame will not be fixed properly.
Affected products
Leksell Gamma Knife Perfexion
Lot or serial number
Not applicable
Model or catalog number
715000
Companies
- Manufacturer
-
Elekta InstrumentS AB
Kungstensgatan 18,
Stockholm
10393
SWEDEN