Leica M220 F12 Microscope System (2018-12-16)

Starting date:
December 16, 2018
Posting date:
January 23, 2019
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-68892

Affected products

Leica  M220 F12 Microscope System

Reason

Leica Microsystems has received two (2) (Nov. 2018) complaints pertaining to the M220 F12 Surgical Microscope System. In both, the optics carrier were reported to have fallen. These complaints detailed mechanical part failures with the screw that connects the optics carrier to the free end of the swing arm. Leica identified the cause of the observed issues to be insufficient securing of the screws that connect the swing arm with the optics carrier.

Affected products

Leica  M220 F12 Microscope System

Lot or serial number

170718002

Model or catalog number

10448367

Companies
Manufacturer
Leica Microsystems (Schweiz) Ag
Max Schmidheiny-Strasse 201
Heerbrugg
9435
SWITZERLAND