Health product recall

Leica Cryostats (2019-03-18)

Starting date:
March 18, 2019
Posting date:
April 12, 2019
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-69560


Last updated:

Affected Products

Leica Cryostats

Reason

During post market surveillance, it was realized that some customers are using flammable freezing sprays within the Cryostat chamber. Such flammable freezing sprays can potentially ignite and can cause serious injuries.  The current device labeling does not include any specific warning about the usage of flammable freezing sprays, therefore the corrective action will improve the device labeling with an adhesive warning label and safety amendment to the instrument instruction for use (IFU).

Affected products

Leica Cryostats

Lot or serial number

More than 100 numbers, contact manufacturer.

Model or catalog number

CM1100

CM1510 S

CM1520

CM1850

CM1850UV

CM1860

CM1860UV

CM1950

CM3050 S

Companies
Manufacturer

LEICA BIOSYSTEMS NUSSLOCH GMBH

HEIDELBERGER STREET 17-19

NUSSLOCH

69226

GERMANY