Leica Cryostats (2019-03-18)
- Starting date:
- March 18, 2019
- Posting date:
- April 12, 2019
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-69560
Last updated: 2019-04-12
Affected Products
Leica Cryostats
Reason
During post market surveillance, it was realized that some customers are using flammable freezing sprays within the Cryostat chamber. Such flammable freezing sprays can potentially ignite and can cause serious injuries. The current device labeling does not include any specific warning about the usage of flammable freezing sprays, therefore the corrective action will improve the device labeling with an adhesive warning label and safety amendment to the instrument instruction for use (IFU).
Affected products
Leica Cryostats
Lot or serial number
More than 100 numbers, contact manufacturer.
Model or catalog number
CM1100
CM1510 S
CM1520
CM1850
CM1850UV
CM1860
CM1860UV
CM1950
CM3050 S
Companies
- Manufacturer
-
LEICA BIOSYSTEMS NUSSLOCH GMBH
HEIDELBERGER STREET 17-19
NUSSLOCH
69226
GERMANY