This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

LEICA ASP6025 TISSUE PROCESSOR(2015-08-11)

Starting date:
August 11, 2015
Posting date:
August 24, 2015
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-54782

Affected Products

A. LEICA ASP6025 TISSUE PROCESSOR

Reason

One complaint was received from a customer in Spain where poor processing was identified and samples were deemed non-diagnosable. the investigation of the unit when returned to the manufacturer on July 15th led to a safety meeting which determined that the instrument is incorrectly getting into a state of "bottle empty" during retort filling, after passing the fill lever sensor 2 and prior to reaching level sensor 3, thus not completing a 5 litre fill.

Affected products

A. LEICA ASP6025 TISSUE PROCESSOR

Lot or serial number

314

316

446

448

538

594

622

Model or catalog number

ASP6025

Companies
Manufacturer

Leica Microsystems Nussloch GMBH

HEIDELBERGER STREET 17-19

Nussloch

69222

GERMANY