This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

LAVA Ultimate Restoratives (2015-06-15)

Starting date:
June 15, 2015
Posting date:
July 7, 2015
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-54090

Recalled Product

  1. LAVA ULTIMATE CAD/CAM RESTORATIVE
  2. LAVA ULTIMATE IMPLANT CROWN RESTORATIVE

Reason

Approximately 10% of dental accounts have reported to 3M that Lava Ultimate Crowns de-bonded at a rate higher than their typical experience for crowns. The reported de-bond rates varied greatly and often significantly exceeded the 2-4% rates generally reported in clinical literature. Based on this customer feedback, the product is not meeting performance expectations for consistent and predictable outcomes for the crown indication.

Affected products

A.LAVA ULTIMATE CAD/CAM RESTORATIVE

Lot or serial number

ALL

Model or catalog number

ALL

Companies
Manufacturer

3M ESPE Dental Products

2510 Conway Avenue

St. Paul

Minnesota

UNITED STATES



B.LAVA ULTIMATE IMPLANT CROWN RESTORATIVE

Lot or serial number

ALL

Model or catalog number

ALL

Companies
Manufacturer

3M ESPE Dental Products

2510 Conway Avenue

St. Paul

Minnesota

UNITED STATES